The Techne® Model 4330 Meter is a dual measurement Conductivity and pH instrument. The instrument contains an enhanced 240 x 128 graphics LCD display. pH can be displayed to 3 decimal places resolution whilst mV can be displayed in 0.1mV resolution. Independent or combined use of sensors is permitted - there is no electrical interference between sensors. Product operation protocols have been carefully designed to ensure ease of use enabling the most complex determinations to be carried out by unfamiliar users with confidence.

Traceability of calibration and sample data is easily maintained and guarantees a high level of confidence. Product performance is continually monitored and resulting messages are clearly displayed. All results and calibrations are date and time stamped. The Model 4330 Conductivity / pH Meter has been designed and manufactured to comply with European CE regulations and the power supply modules meet the specifications of all major European and North American safety standards.

Model 4330 Specifications

Conductivity  
Ranges 0 to 1.999S / 0 to 19.99S with X10 Cell / 0 to 1999g/l / 0 to 20MOhms
Resolution 0.001µS to 0.01S / 0.001mg/l to 1g/l / 0.01MOhms
Accuracy ±0.5% ±2 digits / ±2% ±2digits on MOhm range
pH  
Ranges -2 to 19.999pH / -1999.9 to 1999.9mV
Resolution 0.1/0.01/0.001pH / 1mV/0.1mV
Accuracy ±0.003pH / ±0.2mV
Input Impedance >1013Ohms
Temperature  
Ranges -10 to 105ºC / 14 to 220ºF
Resolution 0.1ºC / 1ºF
Accuracy ±0.5ºC / ±1ºF
ATC Ranges 0 to 100ºC / 32 to 212ºF
Temperature Inputs 2
Salinity  
Range 0 to 99.9
Resolution 0.1
Accuracy 0 to 35 ±1/35 ± 3
Outputs Bi-directional RS232C / Analog output / Karl Fischer sockets / Hi-Lo Alarm outputs open collector / 0.5A 50V DC max / Electrode platinising socket applying 10V AC 0.05Hz.
Alarm Points 0 to 19.99S / 0 to 1999g/l / -2 to 19.999pH / -1999.9 to 1999.9mV
GLP Calibration reminder with an interval of 1 to 999 hours from last calibration. / Timed printout with an interval from 1 second to 1 day. / End point suppress time from 0 seconds to 1 hour.
Display LCD graphics
Power Power supply 9V AC
Size 275(l) x 240(w) x 150(d)mm
Weight 1.2Kg

Model 4330 Order Codes

551 001 Model 4330 supplied with glass conductivity cell with in-built ATC K=1 (027 013), glass pH electrode (924 005), 4 & 7 pH buffer capsules, swing arm electrode holder (903 200), English instruction manual and UK Power Supply (021 030).

A number of additional accessories are available to complement the Model 4330 Conductivity / pH Meter, including pH and Conductivity Electrodes and a 40 column printer.

Phone: 813-971-5001
Fax : 813-971-5117

Omni Controls Inc.
5309 Technology Dr.
Tampa, FL 33647

Toll Free
1-800-783-6664

email sales@omnicontrols.com
Sales@Omnicontrols.com

We guarantee a 24 hour response on all inquiries made via email.
For quicker response please call Omni Controls
at any of the above numbers

 


Please Note: All prices subject to change without notice
copyright©omnicontrols.com - All Rights Reserved.
Terms and Conditions for Omni Controls Inc - online.
Legal Disclaimer
Supplementary Information and Cover Art Copyright omnicontrols.com. All rights reserved.

copyright © omnicontrols 2005
Omni Controls Inc.
5309 Technology Dr.
Tampa, FL 33647

Email: sales@omnicontrols.com

Phone: 800-783-6664
Fax: 813-971-5117