Model 4150 Conductivity Meter
|
 |
The Model 4150 Conductivity Meter, part of the Premier Portable
range, is designed to meet the needs of the user looking for simplicity of
operation combined with accurate and reliable results.
Catalog #: 553-001
List Price $695
|
The Model 4150 Conductivity Meter is supplied with a robust solid epoxy
sensor which eliminates the risk of probe damage during use. This standard
sensor has a nominal cell constant of 1.00. The meter also enables the
digital selection of cell constant values between 0.015 and 19.99 to
facilitate the use of the wide range of cell options offered. The assigned
cell constant values can be checked and updated by the use of standard
solutions and the 4150 is configured to automatically recognize three
common conductivity values across the measuring range (10µs, 1413µs and
12.88µs). Standards, traceable to international standards, are available
as part of the Model 4150 Conductivity Meter package.
The reference temperature is selectable to be 18, 20 or 25°C and the
temperature coefficient can be digitally set to any value between 0 and
4.00%/°C, with a default to 2.00%/°C, a value which is acceptable in many
analytical situations. The Model 4150 Conductivity Meter will also display
results in T.D.S. (Total Dissolved Solids) units using an in-built
conversion factor of 0.66. Other Conductivity Meters
available from Techne® are the Model 4200 Premier Portable
Conductivity Meter, the Model 4310 Laboratory
Conductivity Meter, the Model 4320 Laboratory
Conductivity Meter and the Model 4330 Conductivity / pH
Meter.
Also available from Techne® in the Premier Portable
range are the Model 3150 pH Meter, the
Model 3205 Ion Meter
and the Model 9200
Dissolved Oxygen Meter.

Model 4150 Specifications |
| Conductivity |
|
| Ranges |
5 switched ranges 0 to 199.9mS |
| Resolution |
0.1mS / 0.01mS / 1µS / 0.1µS / 0.01µS |
| Accuracy |
±0.5% ±2 digits |
| TDS |
|
| Ranges |
5 switched ranges 0 to 199.9g/l |
| Resolution |
0.1g/l / 0.01g/l / 1mg/l / 0.1mg/l / 0.01mg/l |
| Accuracy |
±0.5% ±2 digits |
| Temperature |
|
| Ranges |
-10 to +105°C / +14 to +220°F |
| Resolution |
0.1°C / 1°F |
| Accuracy |
±0.5°C / ±1°F |
| ATC Range |
0 to 100°C |
| Cell Constant |
Digitally settable 0.015 to 19.99 |
| Reference Temperature |
18, 20 or 25°C |
| Temperature Coefficient |
0.00 to 4.00/°C |
| Probe |
2 plate configuration |
| Power |
3 AA cells |
| Size |
200(l) x 80(w) x 60(d)mm |
| Weight |
370g |

| 553 001 |
Model 4150 supplied with epoxy conductivity cell K=1 (027
233), batteries and instruction manual |
| 033 174 |
Carrying Case |
| 027 235 |
Cell epoxy bodied with ATC K=0.1 |
| 027 236 |
Cell epoxy bodied with ATC K=10 |
| 027 237 |
Cell glass bodied with ATC K=1 |
| 025 139 |
Calibration Standard 10µS |
| 025 138 |
Calibration Standard 1413µS |
| 025 156 |
Calibration Standard 12.8mS |
(Lead length on probes is normally 1 metre. Details on non-standard
lead lengths are available on request)

|